How to Hook up Multiple Patients to the Same Vent

Today, we’re going to talk about hooking up multiple intubated patients to the same ventilator.  As the coronavirus becomes increasingly more prevalent and more patients require intubation, knowing how to do this may become more important than ever before.

 

  • Why would I do this?

    • As more and more patients require intubation, ventilators are going to become a precious resource.  We want to maximize our resources to help as many people as possible.

    • So what are the downsides?

    • You can no longer adjust the vent to optimize it for a single patient

      • Under normal circumstances, you want to optimize vent settings for a specific patient

      • This allows you to maximize oxygenation and ventilation while also keeping the patient as comfortable as possible on the ventilator in order to avoid having to over-sedate the patient

    • You can no longer allow a patient to trigger the vent

      • Under normal circumstances, vents allow a patient to trigger a breath

        • This is helpful because it is more comfortable for the patient and will allow an intubated patient to be less sedated

      • However, you don’t want one patient triggering breaths on the vent and thereby affecting every other patient

        • For example, if one patient is tachypneic and triggering breaths, all other patients attached to the same vent will be forced to breathe at this rate as well

    • Ventilation is less effective

      • In order to attach multiple patients to a single vent, you will need to use a large amount of tubing with Y-site connectors

      • This results in increased dead space making tidal volumes less accurate

      • As a result, patients are more likely to develop hypercapnia in this setup

        • This will likely require permissive hypercapnia in these patients

  • Now that we understand the downsides to this setup, let’s discuss the settings you will need to consider

    • You will want to use pressure control for these patients

      • Why not volume control?

        • Normally, volume control is helpful because it allows you to provide a specific tidal volume to a patient, but is limited because it allows no control over peak pressure

        • When you have multiple patients attached to a single vent, however, you are no longer able to accurately control the tidal volume of any single patient

        • If there is a problem with one patient, for example if there is an obstruction or kinking of the ET tube, than the other patient on the vent will get significantly increased tidal volume as the volume intended for 2 patients enters only 1.  This can result in dangerously high peak pressures and barotrauma.

        • If you want to hook up multiple patients to the same vent with volume control, they would all need to receive the same volume, meaning they should all be a similar size.  This further restricts which patients may be placed on the same vent.

      • Why is pressure control better?

        • Even with multiple patients hooked up to the same vent, you can maintain adequate control over the peak pressures supplied.

        • If there is a problem with one patient, for example there is an obstruction or kinking of one ET tube, it will not affect the other patient.  The patient with the problematic ET tube will receive less tidal volume, but the other patient on the vent will be unaffected.

        • Different sized patients can be hooked up to the same vent, since larger patients have higher compliance and therefore will receive larger breaths

    • You should set the vent to continuous mandatory ventilation

      • You do not want patients to trigger the vent and thereby affect other patients attached

      • Instead, you need continuous mandatory ventilation, in which the vent is set to a fixed rate and the patient cannot trigger the vent

      • If the vent does not have this mode as an option, you can instead max out the ventilator trigger threshold, thereby preventing patients from triggering the vent

        • If this doesn’t work or the patient is fighting the vent, then you may need to consider sedation that also suppresses the respiratory drive such as opiates and propofol

        • If even that is unsuccessful and the patient is still fighting or triggering the vent, you may need to consider paralytics

  • Ok, now we understand the pros and cons, as well as the settings we need. How do we set this up?

    • First, make sure the patients you are attaching to a single ventilator have similar vent requirements; ie don’t attach a patient who needs an FiO2 of 30% and PEEP of 5 to the same machine as a patient who needs an FiO2 of 100% and PEEP of 15

    • Set up the vent settings as discussed above

    • Attach viral filters to prevent cross-contamination between patients

    • Using Y-site connectors, attach the expiratory and inspiratory limbs of the vent to all ET tubes as shown in the diagram

  • Keep in mind, that this setup has been tested primarily with lung models and animals; there isn’t significant data from human studies.  But it may be important despite this going forward!

Pics courtesy of emcrit.org 

Stay safe everyone!

Vent splitting pic 2.png
vent splitting pic 1.png

Airway Management in a Coronavirus Patient

Today we’re going to forego trauma Tuesday to talk about everyone’s favorite topics nowadays: coronavirus and intubating!

  • Your patient has suspected or known COVID-19 and is starting to desaturate on room air.  Now what?

    • Just like any other patient, the first thing to try is oxygen, either via nasal cannula or NRB

    • You can crank up the nasal cannula as high as 6 in order to help maintain oxygenation

  • Great, but my patient is continuing to desaturate even with oxygen.

    • This is where things change from any other patient:

    • DO NOT USE BIPAP OR HIGH FLOW NC

      • When these patients get very ill, these modalities have a high likelihood of failing them

      • These 2 modalities also will result in significant aerosolized spread of covid-19

        • Even if you put them in a negative pressure isolation room with the bipap, you will have no way of transporting them

      • SO JUST DON’T DO IT

  • OK, so I can’t use bipap or HFNC but my patient is still desaturating…

    • It’s time to intubate!

    • You should intubate early with these patients, and avoid crash intubations whenever possible

    • Step 1: gown up

      • This means gown, gloves, N95, and a mask with face shield over your N95

    • Step 2: pre-oxygenate

      • Pre-oxygenate using NRB

      • You do not want to use apneic oxygenation via nasal cannula, as this will further aerosolize the virus and has marginal evidence supporting it even in the best conditions

      • Do not bag the patient if it can at all be avoided; again, this will aerosolize the virus and result in increased risk of exposure for everyone in the area

    • Step 3: intubate

      • Use VL instead of DL

        • VL allows you to stay farther away from the patients mouth and secretions, helping protect you against the virus

      • The most experienced person should be performing the intubation – you want to maximize the chances for first pass success

    • Step 4: set the vent (or have someone else do it if you’re gowned up)

      • Treat these patients as ARDS patients and use the ARDSnet protocol with low tidal volumes

      • Unlike ARDS, however, steroids do not play a role in management

    • Step 5: de-gown

      • Ideally, have a spotter present so they can help make sure you don’t accidentally contaminate yourself during this process

      • In particular, be careful not to contaminate any mucous membranes, meaning be particularly careful around your eyes, nose, and mouth

      • Wash your hands!

  • Congratulations! You have successfully intubated this patient without unnecessarily exposing yourself or your colleagues to coronavirus!